Sector 20-BM Energy Resolution.

The V K-edge of a V2O5 powder (powder-on-tape) was measured with end-station entrance slits of 0.6 x 6.0 mm (V x H).

The data was processed by fitting the pre-edge peak to a set of Lorenztian lineshapes. The Ifeffit fitting macro is: attachment:v_edge_fit.iff, and the data file is V205_20BM.xmu

Fit results gave the following values for Peak 1, the largest pre-edge peak:

Peak Position

5469.20 (fixed)

Peak Width

1.865 +/- 0.018

Peak Height

1.228 +/- 0.013

Images of the fits here:

v_edge_components.png

v_edge_residual.png